Possibilities and limits of X-ray microtomography for in vivo and ex vivo detection of vascular calcifications

In the present paper, vascular calcifications due to chronic renal failure in rats are studied by X-ray microtomography (μ-CT). Although μ-CT is traditionally used as an imaging technique, a quantitative analysis of data obtained by in vivo and ex vivo μ-CT is described and discussed. By comparison...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:International Journal of Cardiovascular Imaging 2009-08, Vol.25 (6), p.615-624
Hauptverfasser: Postnov, A. A., D’Haese, P. C., Neven, E., De Clerck, N. M., Persy, V. P.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:In the present paper, vascular calcifications due to chronic renal failure in rats are studied by X-ray microtomography (μ-CT). Although μ-CT is traditionally used as an imaging technique, a quantitative analysis of data obtained by in vivo and ex vivo μ-CT is described and discussed. By comparison with traditional destructive methods, such as histomorphometry and atomic absorption, the detection limits for calcium were determined in living rats and in extracted aortas. μ-CT proved to be an effective non-invasive imaging technique allowing non-destructive quantification of ectopic calcifications.
ISSN:1569-5794
1573-0743
1875-8312
DOI:10.1007/s10554-009-9459-9