Image quality for radiotherapy CT simulators with different scanner bore size
•Bore size in radiotherapy simulators.•Same imaging dose image quality comparison.•Tarde-off between imaging dose and radiotherapy precision. We compare image quality parameters derived from phantom images taken on three commercially available radiotherapy CT simulators. To make an unbiased evaluati...
Gespeichert in:
Veröffentlicht in: | Physica medica 2018-01, Vol.45, p.65-71 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | •Bore size in radiotherapy simulators.•Same imaging dose image quality comparison.•Tarde-off between imaging dose and radiotherapy precision.
We compare image quality parameters derived from phantom images taken on three commercially available radiotherapy CT simulators. To make an unbiased evaluation, we assured images were obtained with the same surface dose measured using XR-QA2 model GafChromic™ film placed at the imaging phantom surface for all three CT-simulators.
Radiotherapy CT simulators GE LS 16, Philips Brilliance Big Bore, and Toshiba Aquilion LB were compared in terms of spatial resolution, low contrast detectability, image uniformity, and contrast to noise ratio using CATPHAN-504 phantom, scanned with Head and Pelvis protocols. Dose was measured at phantom surface, with CT scans repeated until doses on all scanners were within 2%.
In terms of spatial resolution, the GE simulator appears slightly better, while Philips CT images are superior in terms of SNR for both scanning protocols. The CNR results show that Philips CT images appear to be better, except for high Z material, while Toshiba appears to fit in between the two simulators.
While the image quality parameters for three RT CT simulators show comparable results, the scanner bore size is of vital importance in various radiotherapy applications. Since the image quality is a function of a large number of confounding parameters, any loss in image quality due to scanner bore size could be compensated by the appropriate choice of scanning parameters, including the exposure and by balancing between the additional imaging dose to the patient and high image quality required in highly conformal RT techniques. |
---|---|