Crossover Evaluation of Compressors and Nebulizers Typically Used by Cystic Fibrosis Patients
Compressor/nebulizer units are used to deliver inhaled medications to patients with cystic fibrosis. Practitioners and parents frequently replace either the compressor or the nebulizer with a similar component from a different brand. We hypothesized that these changes could affect the compressor/neb...
Gespeichert in:
Veröffentlicht in: | Respiratory care 2018-03, Vol.63 (3), p.294-300 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Compressor/nebulizer units are used to deliver inhaled medications to patients with cystic fibrosis. Practitioners and parents frequently replace either the compressor or the nebulizer with a similar component from a different brand. We hypothesized that these changes could affect the compressor/nebulizer flow-pressure and aerosol characteristics.
The following compressors were studied: Pari Vios, Pulmo-Aide model 5650D, and Inspiration Elite model HS456. The following nebulizers were studied: Pari LC Plus, Viox, and SideStream Plus. Units that underwent intense use were tested. The recommended compressor/nebulizer combinations by the manufacturers were compared to all other combinations. In-line measurements of maximal flow and pressure were done for all combinations. A Next Generation Impactor was used to determine particle-size characteristics of albuterol (2.5 mg/3 mL). A breathing simulator programmed to deliver an adult breathing pattern was used. Albuterol concentration was measured with spectrophotometry at 276 nm. The following variables were studied: maximal flow and pressure generated by the compressor/nebulizer, mass median aerodynamic diameter, percentage of drug mass contained in particles < 5 μm, and inhaled mass in the respirable range.
Replacing the nebulizer resulted in changes in the flow-pressure characteristics, particle size, and inhaled mass in the respirable range of the paired compressor/nebulizers. The changes were more pronounced when the nebulizer was replaced than when the compressor was changed.
Our findings indicate that, in general, replacing the nebulizer or compressor with a different brand changes the flow-pressure and aerosol characteristics. Practitioners should be cautious when changing compressor/nebulizer pairs unless they are aware of the resulting impact on the flow-pressure and aerosol characteristics. |
---|---|
ISSN: | 0020-1324 1943-3654 |
DOI: | 10.4187/respcare.05892 |