Liposome polymerase chain reaction assay for the sub-attomolar detection of cholera toxin and botulinum neurotoxin type A
We describe an ultrasensitive immunoassay for detecting biotoxins that uses a liposome with encapsulated DNA reporters, and ganglioside receptors embedded in the bilayer, as the detection reagent. After immobilization of the target biotoxin by a capture antibody and co-binding of the detection reage...
Gespeichert in:
Veröffentlicht in: | Nature protocols 2006-11, Vol.1 (4), p.2003-2011 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We describe an ultrasensitive immunoassay for detecting biotoxins that uses a liposome with encapsulated DNA reporters, and ganglioside receptors embedded in the bilayer, as the detection reagent. After immobilization of the target biotoxin by a capture antibody and co-binding of the detection reagent, the liposomes are ruptured to release the reporters, which are quantified by real-time polymerase chain reaction. The new assays for cholera and botulinum toxins are several orders of magnitude more sensitive than current detection methods. A single 96-well microtiter plate can analyze ∼20 specimens, including calibration standards and controls, with all measurements conducted in triplicate. Using pre-coated and blocked microtiter plates, and pre-prepared liposome reagents, a liposome polymerase chain reaction assay can be carried out in about 6 h. |
---|---|
ISSN: | 1754-2189 1750-2799 |
DOI: | 10.1038/nprot.2006.331 |