T1‐based sensing of mammographic density using single‐sided portable NMR
Purpose A single‐sided NMR instrument was used to investigate the ability of the T1 relaxation constant to distinguish between regions of low and high mammographic density in human breast tissue. Methods Measurements were performed on 5 breast slices obtained from 3 women undergoing breast reduction...
Gespeichert in:
Veröffentlicht in: | Magnetic resonance in medicine 2018-09, Vol.80 (3), p.1243-1251 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose
A single‐sided NMR instrument was used to investigate the ability of the T1 relaxation constant to distinguish between regions of low and high mammographic density in human breast tissue.
Methods
Measurements were performed on 5 breast slices obtained from 3 women undergoing breast reduction surgery or prophylactic mastectomy.
Results
T1 values measured in regions of high mammographic density in both the full breast slices (T1 = 170 ± 30 ms) and excised regions (T1 = 160 ± 30 ms) were found to be significantly different (P |
---|---|
ISSN: | 0740-3194 1522-2594 |