T1‐based sensing of mammographic density using single‐sided portable NMR

Purpose A single‐sided NMR instrument was used to investigate the ability of the T1 relaxation constant to distinguish between regions of low and high mammographic density in human breast tissue. Methods Measurements were performed on 5 breast slices obtained from 3 women undergoing breast reduction...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Magnetic resonance in medicine 2018-09, Vol.80 (3), p.1243-1251
Hauptverfasser: Tourell, Monique C., Ali, Tonima S., Hugo, Honor J, Pyke, Chris, Yang, Samuel, Lloyd, Thomas, Thompson, Erik W., Momot, Konstantin I.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Purpose A single‐sided NMR instrument was used to investigate the ability of the T1 relaxation constant to distinguish between regions of low and high mammographic density in human breast tissue. Methods Measurements were performed on 5 breast slices obtained from 3 women undergoing breast reduction surgery or prophylactic mastectomy. Results T1 values measured in regions of high mammographic density in both the full breast slices (T1 = 170 ± 30 ms) and excised regions (T1 = 160 ± 30 ms) were found to be significantly different (P 
ISSN:0740-3194
1522-2594
DOI:10.1002/mrm.27098