Efficiency and measurements agreement between swept-source OCT and low-coherence interferometry biometry systems
Purpose To compare and evaluate the agreement between the measurements obtained with a swept-source optical coherence tomography (OCT)-based biometer, the IOLMaster® 700 (IOLM), and those obtained by an optical biometer based on optical low-coherence interferometry (OLCI), the Aladdin (ALD); To eval...
Gespeichert in:
Veröffentlicht in: | Graefe's archive for clinical and experimental ophthalmology 2018-03, Vol.256 (3), p.559-566 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose
To compare and evaluate the agreement between the measurements obtained with a swept-source optical coherence tomography (OCT)-based biometer, the IOLMaster® 700 (IOLM), and those obtained by an optical biometer based on optical low-coherence interferometry (OLCI), the Aladdin (ALD); To evaluate the ability to perform biometric measurements in those eyes with transparency alterations.
Methods
Fifty-five eyes of 55 subjects were included in this study. Axial length (AL), corneal power (K, in diopters) and its astigmatism, anterior chamber depth (ACD), central corneal thickness (CCT), and lens thickness (LT) measures were obtained within both biometers, Zeiss IOLMaster 700 and Topcon Aladdin. Results were analyzed and compared using the Student’s paired samples
t
-test, Bland–Altman analysis and intraclass correlation coefficient (ICC).
Results
Mean age was 73.12 ± 2.63 (62–89 years). The IOLM mean AL, K, and LT values did not show a statistically significant difference from ALD values and showed excellent agreement and correlation (ICC = 1.000, 0.970, 0.952). IOLM measured a lower mean ACD (−0.036 mm) and higher CCT measurements (9.296 μm). Those results were statistically different (