DEVELOPMENT OF A HIGH-EFFICIENCY PROTON RECOIL TELESCOPE FOR D-T NEUTRON FLUENCE MEASUREMENT
A high-efficiency proton recoil telescope was developed to determine neutron fluences in neutron fields using the 3H(d,n)4He reaction. A 2-mm thick plastic scintillation detector was employed as a radiator to increase the detection efficiency and compensate for the energy loss of the recoil proton w...
Gespeichert in:
Veröffentlicht in: | Radiation protection dosimetry 2018-08, Vol.180 (1-4), p.417-421 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A high-efficiency proton recoil telescope was developed to determine neutron fluences in neutron fields using the 3H(d,n)4He reaction. A 2-mm thick plastic scintillation detector was employed as a radiator to increase the detection efficiency and compensate for the energy loss of the recoil proton within. Two silicon detectors were employed as the ΔE and E detectors. The distance between the radiator and the E detector was varied between 50 and 150 mm. The telescope had detection efficiencies of 3.5 × 10-3 and 7.1 × 10-4 cm2 for distances of 50 and 100 mm, respectively, which were high enough to determine the neutron fluence in 14.8-MeV neutron fields, with a few thousand cm-2 s-1 fluence rate, within a few hours. |
---|---|
ISSN: | 0144-8420 1742-3406 |
DOI: | 10.1093/rpd/ncx278 |