Comparative evaluation of neonatal bilirubin using Radiometer whole blood co-oximetry and plasma bilirubin methods from Roche Diagnostics and Ortho Clinical Diagnostics
Clinically significant variation has been reported within and between plasma and whole blood total bilirubin methods used to identify neonates for whom clinical intervention for hyperbilirubinemia may be required. To evaluate total bilirubin measurements between the Radiometer whole blood co-oximete...
Gespeichert in:
Veröffentlicht in: | Clinical biochemistry 2018-03, Vol.53, p.88-92 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Clinically significant variation has been reported within and between plasma and whole blood total bilirubin methods used to identify neonates for whom clinical intervention for hyperbilirubinemia may be required.
To evaluate total bilirubin measurements between the Radiometer whole blood co-oximeter and plasma bilirubin methods from Roche Diagnostics and Ortho Clinical Diagnostics using neonatal specimens.
Total bilirubin levels were analyzed by whole blood co-oximetry (Radiometer® ABL90). Specimens were centrifuged and plasma analyzed for total bilirubin with a diazo method (Roche Cobas® C-601) and a reflectance spectrophotometric BuBc dry film method (Ortho Clinical Diagnostics VITROS® 350). Results were evaluated by regression, Bland-Altman comparisons and t-tests.
The patient correlation study yielded the following regression equations in μmol/L: a) Radiometer=1.03 Roche – 3.5μmol/L b) Radiometer=0.98 Ortho – 5.7μmol/L c) Roche=0.97 Ortho – 2.4μmol/L. The mean bias over the range of total bilirubin levels examined was −1.0μmol/L for the Radiometer versus the Roche (p≤0.305); −4.4μmol/L for the Radiometer versus Ortho (p≤0.005) and −4.4μmol/L for the Roche versus Ortho (p≤0.002).
Whole blood total bilirubin measurement using the Radiometer ABL90 blood gas analyzer provides accurate and precise results compared to the Roche plasma diazo method. Compared to the reflectance spectrophotometric method, results are precise and had a small but statistically significant bias of −4.4μmol/L.
•Reliability of whole blood bilirubin by co-oximetry assessed in neonatal specimens.•Need for standardization of plasma bilirubin methods•Comparison of whole blood and plasma bilirubin methods |
---|---|
ISSN: | 0009-9120 1873-2933 |
DOI: | 10.1016/j.clinbiochem.2017.12.009 |