Conception and validation of a protocol for reuse of non-irrigated electrophysiology catheters in a Brazilian teaching hospital

Purpose Since the Brazilian current legislation permits the reuse of single-use devices under a validated processing protocol, the main purpose of our study was to develop and validate a method for reusing non-irrigated electrophysiology catheter (EC). Methods Manual and mechanical processing by ult...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of interventional cardiac electrophysiology 2018, Vol.51 (1), p.45-50
Hauptverfasser: Leichsenring, Mirtes Loeschner, Psaltikidis, Eliane Molina, de Oliveira Figueiredo, Márcio Jansen, Moretti, Maria Luiza, Trabasso, Plínio
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Purpose Since the Brazilian current legislation permits the reuse of single-use devices under a validated processing protocol, the main purpose of our study was to develop and validate a method for reusing non-irrigated electrophysiology catheter (EC). Methods Manual and mechanical processing by ultrasonic washer was associated with the use of enzymatic solution and hydrogen peroxide with a final rinse with filtered water. Validation of the cleaning process, as well as catheter integrity, was done by observing the ECs in stereoscopic microscope at ×60 magnification, followed by HemoCheck-S® (HCS) test to monitor the presence of residual blood on their surfaces. Ethylene oxide (EO) was used for sterilization, and the final validations of the processing were performed by assays of sterility, pyrogenicity, and EO residuals. Lastly, a cost-minimization study was performed. Results Cleaning process demonstrated absence of organic material detectable by HCS at the surfaces of the ECs. Assays for sterility were negative, and assays of EO residuals and endotoxins showed levels under established standards. The number of reuses was settled to a maximum of seven uses for the ECs with handle and ten uses for ECS without handle. The cost-minimization study showed an 84% savings, when considering seven reuses. Conclusions Processing of ECs was validated at all stages. Therefore, reuse of ECs under the conditions that we designed was considered safe for patients and cost-effective for our institution.
ISSN:1383-875X
1572-8595
DOI:10.1007/s10840-017-0301-3