Margin Assessment in Renal Surgery Using a Handheld Optical Coherence Tomography Probe
To assess the use of a handheld optical coherence tomography (OCT) probe for the evaluation of intraoperative surgical margins during partial nephrectomy (PN). In an initial feasibility study, a radical nephrectomy specimen with a 9-cm tumor was cut into 19 sections, exposing 0 mm (n = 8), 1 mm (n =...
Gespeichert in:
Veröffentlicht in: | Urology (Ridgewood, N.J.) N.J.), 2018-03, Vol.113, p.241-245 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | To assess the use of a handheld optical coherence tomography (OCT) probe for the evaluation of intraoperative surgical margins during partial nephrectomy (PN).
In an initial feasibility study, a radical nephrectomy specimen with a 9-cm tumor was cut into 19 sections, exposing 0 mm (n = 8), 1 mm (n = 6), and 2 mm (n = 5) gross margins. OCT was used to determine the margin width in each specimen. Second, a prospective ex vivo assessment of 15 PN tumor specimens was performed with OCT to determine margin status and to measure the attenuation coefficient of tumor and renal parenchyma.
Median OCT margin width measurements for sectioned samples were 0 mm, 0.9 mm (range 0.7-2.9 mm), and 2.7 (range 1.65-2.8 mm) for grossly 0 mm (positive), 1 mm, and 2 mm margins, respectively. The difference between measurements from all margin groups was statistically significant (P |
---|---|
ISSN: | 0090-4295 1527-9995 |
DOI: | 10.1016/j.urology.2017.11.029 |