Factors influencing irradiance of locally fabricated phototherapy devices in Jos, north-central Nigeria
Locally fabricated phototherapy devices (LFPDs) are widely used in Nigeria for the treatment of neonatal jaundice. Ours was a cross-sectional observational study of all LFPDs in major hospitals in Jos between January and March 2015. We evaluated a total of 24 LFPDs. The irradiance at the level of th...
Gespeichert in:
Veröffentlicht in: | Tropical doctor 2018-04, Vol.48 (2), p.142-146 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Locally fabricated phototherapy devices (LFPDs) are widely used in Nigeria for the treatment of neonatal jaundice. Ours was a cross-sectional observational study of all LFPDs in major hospitals in Jos between January and March 2015. We evaluated a total of 24 LFPDs. The irradiance at the level of the baby was in the range of 2–23.9 µW/cm2/nm. Fourteen devices had the recommended irradiance of ≥10 µW/cm2/nm and none had irradiance in the intensive range. Decreasing distance from the baby, presence of reflectors and increasing number of flourecent tubes significantly contributed to higher irradiance. A combination of six tubes, presence of reflectors and a distance of 10 cm from the baby produced a mean irradiance of 23.40 µW/cm2/nm. The irradiance of LFPDs varies widely and can be improved by simple modifications. |
---|---|