Improving surgical results in complex nerve anatomy during implantation of selective upper airway stimulation
The following report presents a case of two late embedded hypoglossus branches during implantation of an upper airway stimulation device that caused a mixed activation of the tongue when included in the stimulation cuff. In the end, correct cuff placement could be achieved by careful examination of...
Gespeichert in:
Veröffentlicht in: | Auris, nasus, larynx nasus, larynx, 2018-06, Vol.45 (3), p.653-656 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The following report presents a case of two late embedded hypoglossus branches during implantation of an upper airway stimulation device that caused a mixed activation of the tongue when included in the stimulation cuff. In the end, correct cuff placement could be achieved by careful examination of the hypoglossal nerve anatomy, precise nerve dissection, tongue motion analysis and intraoperative nerve monitoring. |
---|---|
ISSN: | 0385-8146 1879-1476 |
DOI: | 10.1016/j.anl.2017.10.005 |