Optimization of T2-weighted imaging for shoulder magnetic resonance arthrography by synthetic magnetic resonance imaging

Background Synthetic magnetic resonance imaging (MRI) allows reformatting of various synthetic images by adjustment of scanning parameters such as repetition time (TR) and echo time (TE). Optimized MR images can be reformatted from T1, T2, and proton density (PD) values to achieve maximum tissue con...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Acta radiologica (1987) 2018-08, Vol.59 (8), p.959-965
Hauptverfasser: Lee, Seung Hyun, Lee, Young Han, Hahn, Seok, Yang, Jaemoon, Song, Ho-Taek, Suh, Jin-Suck
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Background Synthetic magnetic resonance imaging (MRI) allows reformatting of various synthetic images by adjustment of scanning parameters such as repetition time (TR) and echo time (TE). Optimized MR images can be reformatted from T1, T2, and proton density (PD) values to achieve maximum tissue contrast between joint fluid and adjacent soft tissue. Purpose To demonstrate the method for optimization of TR and TE by synthetic MRI and to validate the optimized images by comparison with conventional shoulder MR arthrography (MRA) images. Material and Methods Thirty-seven shoulder MRA images acquired by synthetic MRI were retrospectively evaluated for PD, T1, and T2 values at the joint fluid and glenoid labrum. Differences in signal intensity between the fluid and labrum were observed between TR of 500–6000 ms and TE of 80–300 ms in T2-weighted (T2W) images. Conventional T2W and synthetic images were analyzed for diagnostic agreement of supraspinatus tendon abnormalities (kappa statistics) and image quality scores (one-way analysis of variance with post-hoc analysis). Results Optimized mean values of TR and TE were 2724.7 ± 1634.7 and 80.1 ± 0.4, respectively. Diagnostic agreement for supraspinatus tendon abnormalities between conventional and synthetic MR images was excellent (κ = 0.882). The mean image quality score of the joint space in optimized synthetic images was significantly higher compared with those in conventional and synthetic images (2.861 ± 0.351 vs. 2.556 ± 0.607 vs. 2.750 ± 0.439; P 
ISSN:0284-1851
1600-0455
DOI:10.1177/0284185117740761