Evaluation of the Roche® Elecsys and the Diasorin® Liaison S100 kits in the management of mild head injury in the emergency room
The aim of this single-center prospective study is to compare two commercially available S100ß kits (the Roche® Elecsys and the Diasorin® Liaison S100 kits) in terms of analytical and clinical performances in a population admitted in the emergency room for mild traumatic brain injury (mTBI). 110 pat...
Gespeichert in:
Veröffentlicht in: | Clinical biochemistry 2018-02, Vol.52, p.123-130 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The aim of this single-center prospective study is to compare two commercially available S100ß kits (the Roche® Elecsys and the Diasorin® Liaison S100 kits) in terms of analytical and clinical performances in a population admitted in the emergency room for mild traumatic brain injury (mTBI).
110 patients were enrolled from September 2014 to May 2015. Blood sample draws were performed within 3h after head trauma and the study population was split into pediatric and adult subpopulations (>18years of age).
Although both kits correlated well, we observed a significant difference in terms of S100ß levels (P value |
---|---|