Freeform optical design for a nonscanning corneal imaging system with a convexly curved image
Most existing techniques that are typically used by specialists to image the cornea are based on point, slit, or annular scanning due to a narrow field of view. The difficulty in achieving a larger field of view comes from the convex shape of the human eyeball. Field curvature for a refractive imagi...
Gespeichert in:
Veröffentlicht in: | Applied optics (2004) 2017-07, Vol.56 (20), p.5630-5638 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Most existing techniques that are typically used by specialists to image the cornea are based on point, slit, or annular scanning due to a narrow field of view. The difficulty in achieving a larger field of view comes from the convex shape of the human eyeball. Field curvature for a refractive imaging system with positive power is typically negative and thus a concave image surface. In order to view the full cornea and sclera with snapshot imaging, we calculate qualified two- and three-mirror solutions from Seidel aberration theory. A three-mirror solution is further optimized as a high-resolution off-axis imaging system using freeform surfaces, which can obtain a full-field tailored image shell without scanning. The lateral resolution on the cornea is about 10 μm with good modulation transfer function (MTF) and spot performance. To ease the assembly, a monolithic design is achieved with slightly lower resolution, leading to a potential mass production solution. |
---|---|