Mechanical Properties of Various Heat-treated Nickel-titanium Rotary Instruments
The purpose of this study was to compare the bending stiffness, cyclic fatigue, and torsional fracture resistances of heat-treated and conventional nickel-titanium rotary instruments. V-Taper 2 (VT2; #25/.08), V-Taper 2H (V2H; #25/.06), Hyflex CM (HCM; #25/.06), HyFlex EDM (HDM; #25/variable taper),...
Gespeichert in:
Veröffentlicht in: | Journal of endodontics 2017-11, Vol.43 (11), p.1872-1877 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The purpose of this study was to compare the bending stiffness, cyclic fatigue, and torsional fracture resistances of heat-treated and conventional nickel-titanium rotary instruments.
V-Taper 2 (VT2; #25/.08), V-Taper 2H (V2H; #25/.06), Hyflex CM (HCM; #25/.06), HyFlex EDM (HDM; #25/variable taper), and ProTaper Next X2 (PTN; #25/variable taper) were tested. The bending stiffness was measured with the customized device (AEndoS), and the files (n = 15) were fixed at 3 mm from the tip and bent at 45° with respect to their long axis. Cyclic fatigue resistance was tested by pecking and rotating instruments (n = 15) in artificial canal with a 7.8-mm radius and 35° angle of curvature until fracture. The ultimate torsional strength and toughness were estimated by using AEndoS. The file tip of 5 mm was fixed between resin blocks and driven clockwise at 20 rpm until fracture. The results were analyzed by using one-way analysis of variance and Duncan post hoc comparison. The fracture surfaces and longitudinal aspect of each group were examined under the scanning electron microscope.
CM-wire instruments had lower bending stiffness than others. HDM showed the highest cyclic fatigue resistance, followed by VTH and HCM (P |
---|---|
ISSN: | 0099-2399 1878-3554 |