A Preliminary Evaluation of Ovine Bladder Mucosal Damage Associated With 2 Different Indwelling Urinary Catheters
To determine whether a new catheter design with a low-profile, open-ended rounded rather than pointed Foley catheter tip can reduce mucosal damage to the bladder of ewes. Six ewes were randomly assigned to 1 of 2 indwelling urinary catheters—a 16 Fr Foley catheter or a 16 Fr open-tip CystoSure cathe...
Gespeichert in:
Veröffentlicht in: | Urology (Ridgewood, N.J.) N.J.), 2017-12, Vol.110, p.248-252 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | To determine whether a new catheter design with a low-profile, open-ended rounded rather than pointed Foley catheter tip can reduce mucosal damage to the bladder of ewes.
Six ewes were randomly assigned to 1 of 2 indwelling urinary catheters—a 16 Fr Foley catheter or a 16 Fr open-tip CystoSure catheter. After 96 hours, all the animals were sacrificed and their bladder and urethra harvested for analysis.
Image analysis of the bladder surfaces demonstrated a significant decrease in the percentage of bladder area covered by ulceration and inflammation in sheep with CystoSure catheters compared with Foley catheters (P |
---|---|
ISSN: | 0090-4295 1527-9995 |
DOI: | 10.1016/j.urology.2017.08.020 |