Agreement of Corneal Diameter Measurements Obtained by a Swept-source Biometer and a Scheimpflug-based Topographer

PURPOSE:To assess the agreement between a swept-source optical biometry device (IOLMaster 700) and a Scheimpflug-based topography device (Pentacam HR) in measuring the corneal diameter. METHODS:A total of 100 eyes from 100 cataract surgery candidates were enrolled. The white-to-white (WTW) distance...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Cornea 2017-11, Vol.36 (11), p.1373-1376
Hauptverfasser: Salouti, Ramin, Nowroozzadeh, Mohammad Hossein, Tajbakhsh, Zahra, Bagheri, Mansoureh, Ghoreyshi, Maryam, Azizzadeh, Monir, Razeghinejad, Mohammad Reza
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:PURPOSE:To assess the agreement between a swept-source optical biometry device (IOLMaster 700) and a Scheimpflug-based topography device (Pentacam HR) in measuring the corneal diameter. METHODS:A total of 100 eyes from 100 cataract surgery candidates were enrolled. The white-to-white (WTW) distance measurements provided by both devices were recorded and analyzed. A paired t test was performed to compare the mean values of the WTW distance between the IOLMaster 700 and Pentacam HR. The Bland–Altman analysis was used to evaluate the agreement between the measurements of both devices. RESULTS:Mean age (±SD) of the participants was 65.9 ± 3.3 years, and 58 (58%) were men. The mean WTW distances measured by the IOLMaster 700 and Pentacam HR were 11.72 ± 0.45 mm and 11.41 ± 0.42 mm, respectively (P < 0.001). The overall 95% limits of agreement were −0.17 and 0.78 mm. The 95% limits of agreement between the measurements of the 2 devices were −0.30 and 0.84 versus −0.07 and 0.72 for eyes with WTW distances of ≤11.50 mm versus >11.50 mm, respectively. CONCLUSIONS:According to the results of this study, the IOLMaster 700 may overestimate the WTW distance measurements by up to 0.78 mm compared with the Pentacam HR, so these 2 devices should not be used interchangeably for this purpose. The agreement is somehow weaker for eyes with WTW distances of 11.50 mm or less than those with WTW distances greater than 11.50 mm.
ISSN:0277-3740
1536-4798
DOI:10.1097/ICO.0000000000001300