Experimental validation of predicted application accuracies for computer-assisted (CAS) intraoperative navigation with paired-point registration
Purpose The target registration error (TRE) is a crucial parameter to estimate the potential usefulness of computer-assisted navigation intraoperatively. Both image-to-patient registration on base of rigid-body registration and TRE prediction methods are available for spatially isotropic and anisotr...
Gespeichert in:
Veröffentlicht in: | International journal for computer assisted radiology and surgery 2018-03, Vol.13 (3), p.425-441 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose
The target registration error (TRE) is a crucial parameter to estimate the potential usefulness of computer-assisted navigation intraoperatively. Both image-to-patient registration on base of rigid-body registration and TRE prediction methods are available for spatially isotropic and anisotropic data. This study presents a thorough validation of data obtained in an experimental operating room setting with CT images.
Methods
Optical tracking was used to register a plastic skull, an anatomic specimen, and a volunteer to their respective CT images. Plastic skull and anatomic specimen had implanted bone fiducials for registration; the volunteer was registered with anatomic landmarks. Fiducial localization error, fiducial registration error, and total target error (TTE) were measured; the TTE was compared to isotropic and anisotropic error prediction models. Numerical simulations of the experiment were done additionally.
Results
The user localization error and the TTE were measured and calculated using predictions, both leading to results as expected for anatomic landmarks and screws used as fiducials. TRE/TTE is submillimetric for the plastic skull and the anatomic specimen. In the experimental data a medium correlation was found between TRE and target localization error (TLE). Most of the predictions of the application accuracy (TRE) fall in the 68% confidence interval of the measured TTE. For the numerically simulated data, a prediction of TTE was not possible; TRE and TTE show a negligible correlation.
Conclusion
Experimental application accuracy of computer-assisted navigation could be predicted satisfactorily with adequate models in an experimental setup with paired-point registration of CT images to a patient. The experimental findings suggest that it is possible to run navigation and prediction of navigation application accuracy basically defined by the spatial resolution/precision of the 3D tracker used. |
---|---|
ISSN: | 1861-6410 1861-6429 |
DOI: | 10.1007/s11548-017-1653-y |