Performance characteristics of a high-sensitivity cardiac troponin assay using plasma and whole blood samples
•Using AACC Universal Sample Bank, Pylon assay shown to be second-generation hs-cTn.•>90% of whole blood samples of 505 healthy donors were measurable.•Good correlation/concordance with Abbott hsTnI using suspected AMI patient samples•Assay can measure cTnI in whole blood because red blood cells...
Gespeichert in:
Veröffentlicht in: | Clinical biochemistry 2017-12, Vol.50 (18), p.1249-1252 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | •Using AACC Universal Sample Bank, Pylon assay shown to be second-generation hs-cTn.•>90% of whole blood samples of 505 healthy donors were measurable.•Good correlation/concordance with Abbott hsTnI using suspected AMI patient samples•Assay can measure cTnI in whole blood because red blood cells do not interact with the Pylon probe. |
---|---|
ISSN: | 0009-9120 1873-2933 |