Routine Ultrasound Quality Assurance in a Multi-Unit Radiology Department: A Retrospective Evaluation of Transducer Failures
Abstract The importance of quality assurance (QA) in medical ultrasound (US) has been widely recognized and recommendations concerning technical QA have been published over the years. However, the demonstrated impact of a properly working QA protocol on clinical routine has been scarce. We investiga...
Gespeichert in:
Veröffentlicht in: | Ultrasound in medicine & biology 2017-09, Vol.43 (9), p.1930-1937 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Abstract The importance of quality assurance (QA) in medical ultrasound (US) has been widely recognized and recommendations concerning technical QA have been published over the years. However, the demonstrated impact of a properly working QA protocol on clinical routine has been scarce. We investigated the transducer write-off causes for a 5-y period in a multi-unit radiology department with an annual average of 230 transducers in demanding diagnostic use. The transducer faults and the initial observers of the faults leading to transducer write-offs were traced and categorized. The most common cause of transducer write-off was an image uniformity problem or element failure. Mechanical faults or excessive leakage current and defects in the lens constituted smaller yet substantial shares. Our results suggest that a properly working routine QA program can detect majority of the faults before they are reported by users. |
---|---|
ISSN: | 0301-5629 1879-291X |