Liquid electrode plasma-optical emission spectrometry combined with solid-phase preconcentration for on-site analysis of lead
•A new rapid method for the quantitative onsite analysis of lead using LEP-OES.•Solid-phase extraction is combined with LEP-OES to increase selectivity.•The recovery rate was >95% for the analysis of CRMs and ‘real’ samples.•The LEP-OES analysis results were in a good agreement with that of ICP-O...
Gespeichert in:
Veröffentlicht in: | Journal of chromatography. B, Analytical technologies in the biomedical and life sciences Analytical technologies in the biomedical and life sciences, 2017-08, Vol.1060, p.190-199 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | •A new rapid method for the quantitative onsite analysis of lead using LEP-OES.•Solid-phase extraction is combined with LEP-OES to increase selectivity.•The recovery rate was >95% for the analysis of CRMs and ‘real’ samples.•The LEP-OES analysis results were in a good agreement with that of ICP-OES.•Fast analysis of lead at parts-per-billion levels and below permissible range.
A relatively rapid and precise method is presented for the determination of lead in aqueous matrix. The method consists of analyte quantitation using the liquid electrode plasma-optical emission spectrometry (LEP-OES) coupled with selective separation/preconcentration by solid-phase extraction (SPE). The impact of operating variables on the retention of lead in SPEs such as pH, flow rate of the sample solution; type, volume, flow rate of the eluent; and matrix effects were investigated. Selective SPE-separation/preconcentration minimized the interfering effect due to manganese in solution and limitations in lead-detection in low-concentration samples by LEP-OES. The LEP-OES operating parameters such as the electrical conductivity of sample solution; applied voltage; on-time, off-time, pulse count for applied voltage; number of measurements; and matrix effects have also been optimized to obtain a distinct peak for the lead at λmax=405.8nm. The limit of detection (3σ) and the limit of quantification (10σ) for lead determination using the technique were found as 1.9 and 6.5ng mL−1, respectively. The precision, as relative standard deviation, was lower than 5% at 0.1μg mL−1 Pb, and the preconcentration factor was found to be 187. The proposed method was applied to the analysis of lead contents in the natural aqueous matrix (recovery rate:>95%). The method accuracy was verified using certified reference material of wastewaters: SPS-WW1 and ERM-CA713. The results from LEP-OES were in good agreement with inductively coupled plasma optical emission spectrometry measurements of the same samples. The application of the method is rapid (≤5min, without preconcentration) with a reliable detection limit at trace levels. |
---|---|
ISSN: | 1570-0232 1873-376X |
DOI: | 10.1016/j.jchromb.2017.06.016 |