A European Multicientre Study on the Comparison of HIV-1 Viral Loads Between VERIS HIV-1 Assay and Roche COBAS® TAQMAN® HIV-1 test, Abbott RealTime HIV-1 Assay, and Siemens VERSANT HIV-1 Assay

Highlights • Extraction/amplification for HIV-1 RNA quantitation are integrated in one instrument. • Method comparison of HIV-1 viral load results demonstrated low bias below 1,000 cp/mL. • Patient results over time showed similar profiles and trending between methods. • Differing definitions of vir...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of clinical virology 2017-07, Vol.92, p.75-82
Hauptverfasser: Braun, Patrick, Delgado, Rafael, Drago, Monica, Fanti, Diana, Fleury, Hervé, Hofmann, Jörg, Izopet, Jacques, Kühn, Sebastian, Lombardi, Alessandra, Mancon, Alessandro, Angeles Marcos, M.a, Mileto, Davide, Sauné, Karine, O'Shea, Siobhan, Pérez-Rivilla, Alfredo, Ramble, John, Trimoulet, Pascale, Vila, Jordi, Whittaker, Duncan, Artus, Alain, Rhodes, Daniel
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Highlights • Extraction/amplification for HIV-1 RNA quantitation are integrated in one instrument. • Method comparison of HIV-1 viral load results demonstrated low bias below 1,000 cp/mL. • Patient results over time showed similar profiles and trending between methods. • Differing definitions of virologic failure level may cause confusion with patient management. • Care should be taken when changing systems as variability exists in assay results.
ISSN:1386-6532
1873-5967
DOI:10.1016/j.jcv.2017.05.003