Quantification of breast stiffness using MR elastography at 3 Tesla with a soft sternal driver: A reproducibility study

Purpose Previous studies of breast MR elastography (MRE) evaluated the technique at magnetic field strengths of 1.5 Tesla (T) with the breast in contact with the driver. The aim of this study is to evaluate breast stiffness measurements and their reproducibility using a soft sternal driver at 3T and...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of magnetic resonance imaging 2017-05, Vol.45 (5), p.1379-1384
Hauptverfasser: Hawley, Jeffrey R., Kalra, Prateek, Mo, Xiaokui, Raterman, Brian, Yee, Lisa D., Kolipaka, Arunark
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Purpose Previous studies of breast MR elastography (MRE) evaluated the technique at magnetic field strengths of 1.5 Tesla (T) with the breast in contact with the driver. The aim of this study is to evaluate breast stiffness measurements and their reproducibility using a soft sternal driver at 3T and compare the results with qualitative measures of breast density. Materials and Methods Twenty‐two healthy volunteers each underwent two separate breast MRE scans in a 3T MRI. MRE vibrations were introduced into the breasts at 60 Hz using a soft sternal driver and axial slices were collected using a gradient echo MRE sequence. Mean stiffness measurements were calculated for each volunteer as well as a measure of reproducibility using concordance correlation between scans. Mean stiffness values for each volunteer were assessed and related to amounts of fibroglandular tissue (i.e., breast lobules, ducts, and fibrous connective tissue). Results The stiffness values were reproducible with a significant P‐value 
ISSN:1053-1807
1522-2586
DOI:10.1002/jmri.25511