Complications with the MICRA TPS Pacemaker System: Persistent Complete Heart Block and Late Capture Failure

A Medtronic MICRA transcatheter pacing system (Medtronic, Minneapolis, MN, USA) was implanted in an 86‐year‐old patient with sick sinus syndrome and left bundle branch block after transfemoral aortic valve implantation. During implantation she developed a persistent complete heart block due to manip...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Pacing and clinical electrophysiology 2017-04, Vol.40 (4), p.455-456
Hauptverfasser: HOLM, NIELS, MÜLLER, ANDREAS, ZBINDEN, RAINER
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A Medtronic MICRA transcatheter pacing system (Medtronic, Minneapolis, MN, USA) was implanted in an 86‐year‐old patient with sick sinus syndrome and left bundle branch block after transfemoral aortic valve implantation. During implantation she developed a persistent complete heart block due to manipulation with the large‐bore delivery catheter. Two weeks later, acute pacemaker dysfunction occurred due to massive increase of pacing threshold and impedance without obvious pacemaker dislocation or myocardial perforation. Recurrent capture failure was seen with pacing output set at 5 V/1.0 ms. Hence, microdislocation or fixation of the tines in the right ventricular trabeculae has to be assumed.
ISSN:0147-8389
1540-8159
DOI:10.1111/pace.12998