Performance testing of the neutron flux monitors from 10keV to 1MeV developed for BNCT: A preliminary study

The neutron flux monitors from 10keV to 1MeV designed for boron neutron capture therapy (BNCT) were experimentally tested with prototype monitors in an appropriate neutron field produced at the intense deuterium-tritium neutron source facility OKTAVIAN of Osaka University, Japan. The experimental te...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Applied radiation and isotopes 2017-07, Vol.125, p.119-123
Hauptverfasser: Guan, Xingcai, Manabe, Masanobu, Tamaki, Shingo, Sato, Fuminobu, Murata, Isao, Wang, Tieshan
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The neutron flux monitors from 10keV to 1MeV designed for boron neutron capture therapy (BNCT) were experimentally tested with prototype monitors in an appropriate neutron field produced at the intense deuterium-tritium neutron source facility OKTAVIAN of Osaka University, Japan. The experimental test results and related analysis indicated that the performance of the monitors was good and the neutron fluxes from 10keV to 1MeV of practical BNCT neutron sources can be measured within 10% by the monitors. •Neutron flux monitors from 10keV to 1MeV were developed for BNCT.•The performance of the prototype monitors was experimentally studied.•The monitors worked well and they will be efficiently applicable in BNCT.
ISSN:0969-8043
1872-9800
DOI:10.1016/j.apradiso.2017.04.014