Iridium-Knife: Another knife in radiation oncology

Abstract Purpose Intratarget dose escalation with superior conformity is a defining feature of three-dimensional (3D) iridium-192 (192 Ir) high-dose-rate (HDR) brachytherapy (BRT). In this study, we analyzed the dosimetric characteristics of interstitial192 Ir HDR BRT for intrathoracic and cerebral...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Brachytherapy 2017-07, Vol.16 (4), p.884-892
Hauptverfasser: Milickovic, Natasa, Tselis, Nikolaos, Karagiannis, Efstratios, Ferentinos, Konstantinos, Zamboglou, Nikolaos
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Abstract Purpose Intratarget dose escalation with superior conformity is a defining feature of three-dimensional (3D) iridium-192 (192 Ir) high-dose-rate (HDR) brachytherapy (BRT). In this study, we analyzed the dosimetric characteristics of interstitial192 Ir HDR BRT for intrathoracic and cerebral malignancies. We examined the dose gradient sharpness of HDR BRT compared with that of linear accelerator–based stereotactic radiosurgery and stereotactic body radiation therapy, usually called X-Knife, to demonstrate that it may as well be called a Knife. Methods and Materials Treatment plans for 10 patients with recurrent glioblastoma multiforme or intrathoracic malignancies, five of each entity, treated with X-Knife (stereotactic radiosurgery for glioblastoma multiforme and stereotactic body radiation therapy for intrathoracic malignancies) were replanned for simulated HDR BRT. For 3D BRT planning, we used identical structure sets and dose prescription as for the X-Knife planning. The indices for qualitative treatment plan analysis encompassed planning target volume coverage, conformity, dose falloff gradient, and the maximum dose–volume limits to different organs at risk. Results Volume coverage in HDR plans was comparable to that calculated for X-Knife plans with no statistically significant difference in terms of conformity. The dose falloff gradient—sharpness—of the HDR plans was considerably steeper compared with the X-Knife plans. Conclusions Both 3D192 Ir HDR BRT and X-Knife are effective means for intratarget dose escalation with HDR BRT achieving at least equal conformity and a steeper dose falloff at the target volume margin. In this sense, it can reasonably be argued that 3D192 Ir HDR BRT deserves also to be called a Knife, namely Iridium-Knife.
ISSN:1538-4721
1873-1449
DOI:10.1016/j.brachy.2017.03.005