Influence of Floating-Mass Transducer Coupling Efficiency for Active Middle-Ear Implants on Speech Recognition
OBJECTIVE:The efficiency of vibroplasty (coupler-floating mass transducer [FMT] assembly) can be monitored by direct stimulation of the inner ear through the active middle-ear implant system and comparison of the vibroplasty in vivo threshold and the postoperative bone-conduction pure-tone threshold...
Gespeichert in:
Veröffentlicht in: | Otology & neurotology 2017-07, Vol.38 (6), p.809-814 |
---|---|
Hauptverfasser: | , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | OBJECTIVE:The efficiency of vibroplasty (coupler-floating mass transducer [FMT] assembly) can be monitored by direct stimulation of the inner ear through the active middle-ear implant system and comparison of the vibroplasty in vivo threshold and the postoperative bone-conduction pure-tone threshold. The aim of this study was to compare the vibroplasty in vivo threshold with the postoperative speech recognition in patients with a high preoperative maximum word recognition score.
STUDY DESIGN:Retrospective cohort study of German-speaking patients implanted with a vibrating ossicular prosthesis (VORP) 502 or VORP 503 and high preoperative maximum word recognition score between the years of 2011 and 2015.
SETTING:Multicenter study of four German centers.
PATIENTS:Twenty-three active middle-ear implant users.
INTERVENTION:Rehabilitative.
MAIN OUTCOME MEASURES:Bone-conduction pure-tone and vibroplasty thresholds, postoperative aided word recognition score (WRS) at 65 dB SPL (sound pressure level) and preoperative maximum WRS with Freiburg monosyllabic words.
RESULTS:The mean postoperative aided WRS at 65 dB SPL was 82%. An increasing difference between vibroplasty thresholds and bone-conduction thresholds was associated with a higher discrepancy between the unaided maximum WRS and the postoperative aided WRS. Only if this difference was less than 20 dB, an articulation index of 0.5 (WRS = 75%) or more was achieved.
CONCLUSIONS:Audiological outcome after vibroplasty depends on the coupling efficiency reflected by the vibroplasty threshold. |
---|---|
ISSN: | 1531-7129 1537-4505 |
DOI: | 10.1097/MAO.0000000000001412 |