Fully automated disc diffusion for rapid antibiotic susceptibility test results: a proof-of-principle study
Antibiotic resistance poses a significant threat to patients suffering from infectious diseases. Early readings of antibiotic susceptibility test (AST) results could be of critical importance to ensure adequate treatment. Disc diffusion is a well-standardized, established and cost-efficient AST proc...
Gespeichert in:
Veröffentlicht in: | Journal of antimicrobial chemotherapy 2017-06, Vol.72 (6), p.1659-1668 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Antibiotic resistance poses a significant threat to patients suffering from infectious diseases. Early readings of antibiotic susceptibility test (AST) results could be of critical importance to ensure adequate treatment. Disc diffusion is a well-standardized, established and cost-efficient AST procedure; however, its use in the clinical laboratory is hampered by the many manual steps involved, and an incubation time of 16-18 h, which is required to achieve reliable test results.
We have evaluated a fully automated system for its potential for early reading of disc diffusion diameters after 6-12 h of incubation. We assessed availability of results, methodological precision, categorical agreement and interpretation errors as compared with an 18 h standard. In total, 1028 clinical strains (291 Escherichia coli , 272 Klebsiella pneumoniae , 176 Staphylococcus aureus and 289 Staphylococcus epidermidis ) were included in this study. Disc diffusion plates were streaked, incubated and imaged using the WASPLab TM automation system.
Our results demonstrate that: (i) early AST reading is possible for important pathogens; (ii) methodological precision is not hampered at early timepoints; and (iii) species-specific reading times must be selected. As inhibition zone diameters change over time and are phenotype/drug combination dependent, specific cut-offs and expert rules will be essential to ensure reliable interpretation and reporting of early susceptibility testing results. |
---|---|
ISSN: | 0305-7453 1460-2091 |