In silico investigation of cornea deformation during irrigation/aspiration in phacoemulsification in cataract surgery

Highlights • Cornea deformation was investigated during Irrigation/Aspiration in Phacoemulsification. • Hyper elastic property of cornea and the thickness variation are included in the model. • Surge in the anterior chamber occurs in a negative gauge pressure. • Anterior chamber instability occurs a...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Medical engineering & physics 2017-05, Vol.43, p.77-85
Hauptverfasser: Bayatpour, Dariush, Abouali, Omid, Ghaffarieh, Alireza, Ahmadi, Goodarz
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Highlights • Cornea deformation was investigated during Irrigation/Aspiration in Phacoemulsification. • Hyper elastic property of cornea and the thickness variation are included in the model. • Surge in the anterior chamber occurs in a negative gauge pressure. • Anterior chamber instability occurs at higher volume flow rates for coaxial system. • Deformation of the cornea is less for bimanual system.
ISSN:1350-4533
1873-4030
DOI:10.1016/j.medengphy.2017.02.010