Comparison of specular microscopy and ultrasound pachymetry before and after cataract surgery
Purpose To compare specular microscopy (Topcon SP-3000P, Topcon Corporation, Tokyo, Japan) and ultrasound (US) technology when evaluating central corneal thickness (CCT) prior to and after phacoemulsification. Methods Corneal edema was assessed in phacoemulsification patients due to senile cataract...
Gespeichert in:
Veröffentlicht in: | Graefe's archive for clinical and experimental ophthalmology 2017-02, Vol.255 (2), p.387-392 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose
To compare specular microscopy (Topcon SP-3000P, Topcon Corporation, Tokyo, Japan) and ultrasound (US) technology when evaluating central corneal thickness (CCT) prior to and after phacoemulsification.
Methods
Corneal edema was assessed in phacoemulsification patients due to senile cataract by measuring CCT preoperatively and 1 day, 1 month, and 3 months postoperatively. Bland–Altman analysis was performed to assess interchangeability between pachymetry techniques for each visit. Repeated measures analysis of variance was performed to evaluate variation in CCT agreement depending on the degree of corneal edema.