Development of a method for rapid analysis of Ra-226 in groundwater and discharge water samples by ICP-QQQ-MS
A novel and rapid method has been developed for pre-concentration and measurement of 226Ra in groundwater and discharge water samples using the latest generation triple quadrupole inductively coupled plasma mass spectrometry (ICP-QQQ-MS). Cation exchange and extraction chromatography are capable of...
Gespeichert in:
Veröffentlicht in: | Applied radiation and isotopes 2017-08, Vol.126, p.31-34 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A novel and rapid method has been developed for pre-concentration and measurement of 226Ra in groundwater and discharge water samples using the latest generation triple quadrupole inductively coupled plasma mass spectrometry (ICP-QQQ-MS). Cation exchange and extraction chromatography are capable of pre-concentration factors of ~200 based on 1L samples. The sensitivity and interference removal capability of ICP-QQQ-MS was assessed from spiked groundwaters, with the introduction of He collision gas required to minimise instrument background in high-matrix samples. The technique developed is potentially capable of detecting 226Ra activities as low as 5mBqL−1 when combined with pre-concentration prior to measurement.
•Development of a novel method for the analysis of 226Ra by ICP-QQQ-MS.•Method is capable of 226Ra analysis in 1day.•ICP-QQQ-MS reduces the count time from ~5 days, to 5min per sample.•Potential detection of 5mBqL−1 from a 1L sample. |
---|---|
ISSN: | 0969-8043 1872-9800 |
DOI: | 10.1016/j.apradiso.2017.02.019 |