Experimental comparison of performances of Mega Acer Kit, Ranger and ThermoSens according to flow rates and distances
We experimentally investigated the fluid warming performances of three warmers with different technology, according to flow rates and distances. We used the following intravenous fluid warmers: Mega Acer Kit (Group M, n = 8), Ranger (group R, n = 8), and ThermoSens (group T, n = 8). Fluids that had...
Gespeichert in:
Veröffentlicht in: | Journal of clinical monitoring and computing 2018-12, Vol.32 (6), p.1127-1134 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We experimentally investigated the fluid warming performances of three warmers with different technology, according to flow rates and distances. We used the following intravenous fluid warmers: Mega Acer Kit (Group M, n = 8), Ranger (group R, n = 8), and ThermoSens (group T, n = 8). Fluids that had been stored in the operating room over the previous 24 h were delivered at sequent flow rates of from 440 mL/h up to 2500 mL/h through preheated warming devices. The fluid temperatures were recorded at the inlet point, 76-cm proximal (P
out1
) and 166-cm distal outlet points (P
out2
) every 1 min for 10 min. We repeated each test eight times. The delivered fluid temperature [mean (95% confidence interval)] was significantly higher in group M than group R and T at flow rates up to 650 mL/h with the highest value at 440 mL/h [34.30 (33.35–35.24)°C] (P |
---|---|
ISSN: | 1387-1307 1573-2614 |
DOI: | 10.1007/s10877-017-9995-0 |