Micro-capillary-based self-referencing surface plasmon resonance biosensor for determination of transferrin

A novel self-referencing surface plasmon resonance (SPR) biosensor for detection of transferrin is demonstrated using a micro-capillary as the sensing element. The biosensor employs the SPR mode as a measuring signal and the Fabry-Perot (FP) mode as a referencing signal. The SPR mode is generated in...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Applied Optics 2016-10, Vol.55 (30), p.8571-8575
Hauptverfasser: Chen, Shimeng, Liu, Yun, Liu, Zigeng, Chu, Shuwen, Peng, Wei
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A novel self-referencing surface plasmon resonance (SPR) biosensor for detection of transferrin is demonstrated using a micro-capillary as the sensing element. The biosensor employs the SPR mode as a measuring signal and the Fabry-Perot (FP) mode as a referencing signal. The SPR mode is generated in the gold film that is coated on the outside of the capillary; instead, the FP mode is excited in the capillary, which is filled with de-ionized water. The FP mode is sensitive to temperature and insensitive to refractive index, which can be used as a referencing signal to compensate the effects caused by the temperature fluctuation. The sensor provides a high sensitivity of 1783.943 nm/RIU (refractive index unit) and a resolution of about 7.287×10   RIU. The self-referencing biosensor was applied to measurement of transferrin protein. It can monitor the interaction of transferrin protein with anti-transferrin in real time (0-5.228 μM). The simple and low-cost SPR sensor can be used for highly sensitive self-referencing biosensing for further investigations.
ISSN:0003-6935
1559-128X
2155-3165
1539-4522