A passive dosimeter for evaluating exposure to hydrogen fluoride

A passive dosimeter designed to measure exposure to hydrogen fluoride was tested under laboratory conditions over the concentration range of 0.56 - 4.47 mg HF/m super(3). The lowest concentration was chosen to be close to the Polish TLV (0.5 mg HF/m super(3)). The dosimeter consists of a Whatman No....

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Fluoride 2002-02, Vol.35 (1), p.22-27
Hauptverfasser: Czarnowski, W, Wielgomas, B, Krechniak, J
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A passive dosimeter designed to measure exposure to hydrogen fluoride was tested under laboratory conditions over the concentration range of 0.56 - 4.47 mg HF/m super(3). The lowest concentration was chosen to be close to the Polish TLV (0.5 mg HF/m super(3)). The dosimeter consists of a Whatman No. 1 filter paper impregnated with K sub(2)HPO sub(4) placed in a polyethylene holder protected by a polyethylene screen. After exposure the filter is eluted with 0.1 M sodium citrate and the fluoride concentration determined potentiometrically.
ISSN:0015-4725