Inclusion of Regional Body Composition Parameters Improves Bone Mineral Density Cross-Calibration Between GE Lunar Prodigy and iDXA Densitometers
Abstract Since 1989, the Kuopio Osteoporosis Risk Factor and Prevention (OSTPRE, n = 14220) Study has followed long-term changes of bone mineral density (BMD) and body composition in women with GE Lunar devices. During the course of OSTPRE, the dual-energy X-ray absorptiometry device had to be repla...
Gespeichert in:
Veröffentlicht in: | Journal of clinical densitometry 2017-01, Vol.20 (1), p.97-105 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Abstract Since 1989, the Kuopio Osteoporosis Risk Factor and Prevention (OSTPRE, n = 14220) Study has followed long-term changes of bone mineral density (BMD) and body composition in women with GE Lunar devices. During the course of OSTPRE, the dual-energy X-ray absorptiometry device had to be replaced by a newer model. Then, it was essential to determine whether systematic measurement differences in BMD and body composition will occur. As a part of the OSTPRE study, BMD was measured in 54 women, whereas body composition was determined in 55 women, aged 27–71, by using both the GE Healthcare Lunar Prodigy and iDXA narrow-angle fan beam densitometers during the same visit. The total body fat mass (FM) and lean body mass (LBM) results of these scanners showed a high linear correlation ( r = 0.981–0.994, p |
---|---|
ISSN: | 1094-6950 1559-0747 |
DOI: | 10.1016/j.jocd.2016.07.006 |