Mathematical Discrepancies of the Tono-Pen Applanation Tonometer

PURPOSE OF THE STUDY:The purpose of the study was to determine if Tono-Pen tonometers use simple average and coefficient of variation (CV) algorithms to calculate intraocular pressure (IOP). MATERIAL AND METHODS:IOPs were measured as part of routine ocular examination in 152 client-owned dogs. Using...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of glaucoma 2017-02, Vol.26 (2), p.e30-e36
Hauptverfasser: Doering, Clinton J, Feldman, Eugenia, Bdolah-Abram, Tali, Merideth, Reuben E, Ofri, Ron
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:PURPOSE OF THE STUDY:The purpose of the study was to determine if Tono-Pen tonometers use simple average and coefficient of variation (CV) algorithms to calculate intraocular pressure (IOP). MATERIAL AND METHODS:IOPs were measured as part of routine ocular examination in 152 client-owned dogs. Using 11 Tono-Pen’s, a total of 778 averaged readings were collected. Individual IOP readings, and average IOP and CV displayed by the instrument, were recorded. Average IOP and CV were then manually calculated from individual readings and compared with those displayed by the instrument. RESULTS:The mean absolute difference between the calculated and displayed average IOP was 1.37±2.01 mm Hg (P20%, but displayed CV was
ISSN:1057-0829
1536-481X
DOI:10.1097/IJG.0000000000000524