Trace analysis of triazines and organophosphorus pesticides in water

A method for the determination of trace levels of triazines and organophosphorus pesticides in water is presented. The extraction method is based on a solid-phase extraction on C-18 bound silica SPE cartridges. A precolumn filled with Merck C-18 bound silica and home-made C-18 bound silica have been...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Analytical and bioanalytical chemistry 1996-06, Vol.355 (5-6), p.748-749
Hauptverfasser: CULEA, M, FENESAN, I, COBZAC, S, GOCAN, S, CHIRIAC, M, PALIBRODA, N
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A method for the determination of trace levels of triazines and organophosphorus pesticides in water is presented. The extraction method is based on a solid-phase extraction on C-18 bound silica SPE cartridges. A precolumn filled with Merck C-18 bound silica and home-made C-18 bound silica have been tested at a flow-rate of 3 ml/min with comparable preconcentration yields. A SIM-MS method using a (15)N labelled internal standard has been developed for the organophosphorus pesticides. Detection limits lower than 1 microg/L have been obtained. Separations have been carried out on a conventional GC column OV 17 (1 m) and a capillary column OV 17 (25 m) with a temperature program from 150 degrees C (2 min) to 300 degrees C (rate of 6 degrees C/min).
ISSN:0937-0633
1618-2642
1432-1130
1618-2650
DOI:10.1007/s0021663550748