Capillary blood sampling: relation between lancet diameter, lancing pain and blood volume
The relation between pain of capillary blood sampling and lancet diameter was studied in 52 healthy subjects. The tips of six fingers were pricked with a Softclix® II lancing device using lancets with identical facet geometry but needle diameters of 0.3, 0.4 and 0.8 mm. Two penetration depths (0.9 a...
Gespeichert in:
Veröffentlicht in: | European journal of pain 1999-09, Vol.3 (3), p.283-286 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The relation between pain of capillary blood sampling and lancet diameter was studied in 52 healthy subjects. The tips of six fingers were pricked with a Softclix® II lancing device using lancets with identical facet geometry but needle diameters of 0.3, 0.4 and 0.8 mm. Two penetration depths (0.9 and 1.2 mm) were applied. Pain intensity and blood volume were recorded.
At a puncture depth of 0.9 mm, pain did not differ between the three lancets. Pain increased with penetration depth, and at 1.2 mm the thicker lancets were somewhat more painful than the thinner ones. Blood volumes increased with lancet diameter and penetration depth. If the pain for punctures providing just enough blood for a glucose test (≥10 μl) was compared, there was no difference between the lancets.
It can be concluded that lancet diameter is of minor importance for the patient. |
---|---|
ISSN: | 1090-3801 1532-2149 |
DOI: | 10.1016/S1090-3801(99)90055-1 |