A Rapid and Simple Liquid-Chromatography-Tandem Mass Spectrometry Method for Measuring 25-Hydroxyvitamin D2 and 25-Hydroxyvitamin D3 in Human Serum: Comparison with Two Automated Immunoassays
Total 25-hydroxyvitamin D (25(OH)D) is well-known to be a reliable biomarker of human vitamin D status, with the recognition of widespread vitamin D insufficiency in general populations. The aims of this study are to validate a fast and simple liquid chromatography-tandem mass spectrometry (LC-MS/MS...
Gespeichert in:
Veröffentlicht in: | Annals of clinical and laboratory science 2016-12, Vol.46 (6), p.645-653 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Total 25-hydroxyvitamin D (25(OH)D) is well-known to be a reliable biomarker of human vitamin D status, with the recognition of widespread vitamin D insufficiency in general populations. The aims of this study are to validate a fast and simple liquid chromatography-tandem mass spectrometry (LC-MS/MS) method for quantifying 25(OH)D
and 25(OH)D
in serum and to compare two automated immunoassays with the LC-MS/MS method.
Samples were prepared by protein precipitation with ethanol including 25(OH)D
-d
, followed by a liquid-liquid extraction with hexane. The analytes were separated within a total run time of 3 min. Accuracy was evaluated with standard reference materials (SRM) 972a. Using 150 samples, the LC-MS/MS method was compared with the LIAISON
assay and ADVIA Centaur
assay.
The LC-MS/MS method had a limit of quantitation of 1 ng/mL for the 25(OH)D
and 25(OH)D
with linear responses between 1 and 100 ng/mL. Intra- and inter-assay precision were |
---|---|