Heterogeneity and scatter effects on Ir-192 brachytherapy dose distribution
Highlights • Monte Carlo simulation and film measurement 0f dose near an Ir-192 source. • We studied the effect different size air cavities and scattering medium on dose. • An air cavity of 7 mm diameter increased the dose by 9% distal to the cavity. • The absence of full scatter reduced the dose by...
Gespeichert in:
Veröffentlicht in: | Physica medica 2016-10, Vol.32 (10), p.1210-1215 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Highlights • Monte Carlo simulation and film measurement 0f dose near an Ir-192 source. • We studied the effect different size air cavities and scattering medium on dose. • An air cavity of 7 mm diameter increased the dose by 9% distal to the cavity. • The absence of full scatter reduced the dose by 7% at 10 cm from Ir-192 source. • Air cavities and absence of full scatter induce unacceptable dose errors. |
---|---|
ISSN: | 1120-1797 1724-191X |
DOI: | 10.1016/j.ejmp.2016.09.004 |