Highlights • Monte Carlo simulation and film measurement 0f dose near an Ir-192 source. • We studied the effect different size air cavities and scattering medium on dose. • An air cavity of 7 mm diameter increased the dose by 9% distal to the cavity. • The absence of full scatter reduced the dose by...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Physica medica 2016-10, Vol.32 (10), p.1210-1215
Hauptverfasser: Osman, Alexander F.I, Maalej, Nabil, Ul-Rahman, Khateeb, Rahman, Wamied Abdel
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Highlights • Monte Carlo simulation and film measurement 0f dose near an Ir-192 source. • We studied the effect different size air cavities and scattering medium on dose. • An air cavity of 7 mm diameter increased the dose by 9% distal to the cavity. • The absence of full scatter reduced the dose by 7% at 10 cm from Ir-192 source. • Air cavities and absence of full scatter induce unacceptable dose errors.