Pulmonary vein reconnection following catheter ablation of atrial fibrillation using the second-generation cryoballoon versus open-irrigated radiofrequency: results of a multicenter analysis
Purpose Catheter ablation of atrial fibrillation (CAAF) using the cryoballoon has emerged as an alternate strategy to point-by-point radiofrequency. However, there is little comparative data on long-term durability of pulmonary vein (PV) isolation comparing these two modalities. Methods In this mult...
Gespeichert in:
Veröffentlicht in: | Journal of interventional cardiac electrophysiology 2016-12, Vol.47 (3), p.341-348 |
---|---|
Hauptverfasser: | , , , , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose
Catheter ablation of atrial fibrillation (CAAF) using the cryoballoon has emerged as an alternate strategy to point-by-point radiofrequency. However, there is little comparative data on long-term durability of pulmonary vein (PV) isolation comparing these two modalities.
Methods
In this multicenter, retrospective analysis, the incidences/patterns of late PV reconnection following an index CAAF using the second-generation cryoballoon versus open-irrigated, non-force-sensing radiofrequency were examined.
Results
Of the 2002 patients who underwent a first-time CAAF, 186/1126 patients (16.5 %) ablated using cryoballoon and 174/876 patients (19.9 %) with non-contact force-guided radiofrequency required a repeat procedure at 11 ± 5 months. During follow-up, the incidence of atrial flutters/tachycardias was lower (19.9 vs. 32.8 %;
p
= 0.005) and fewer patients exhibited PV reconnection (47.3 vs. 60.9 %;
p
= 0.007) with cryoballoon versus radiofrequency. Additionally, fewer PVs had reconnected with cryoballoon versus radiofrequency (18.8 vs. 34.6 %;
p
|
---|---|
ISSN: | 1383-875X 1572-8595 |
DOI: | 10.1007/s10840-016-0172-z |