Assessing residual contamination and damage inside flexible endoscopes over time

Highlights • Gastroscopes had higher ATP levels than colonoscopes. • Borescope examinations of endoscope lumens revealed defects requiring repair. • Microbial growth was found on approximately half of patient-ready endoscopes. • Internal damage and residual fluid may foster contamination and biofilm...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:American journal of infection control 2016-12, Vol.44 (12), p.1675-1677
Hauptverfasser: Ofstead, Cori L., MSPH, Wetzler, Harry P., MD, MSPH, Eiland, John E., MS, RN, Heymann, Otis L., BA, Held, Sarah B., MBA, RN, Shaw, Michael J., MD
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Highlights • Gastroscopes had higher ATP levels than colonoscopes. • Borescope examinations of endoscope lumens revealed defects requiring repair. • Microbial growth was found on approximately half of patient-ready endoscopes. • Internal damage and residual fluid may foster contamination and biofilm formation.