Two loose screws: near-miss fall of a morbidly obese patient after an operating room table failure
Abstract Operating room surgical table failure is a rare event but can lead to a dangerous situation when it does occur. The dangers can be compounded in the presence of obesity, especially in the anesthetized or sedated patient. We present a case of a near-miss fall of a morbidly obese patient whil...
Gespeichert in:
Veröffentlicht in: | Journal of clinical anesthesia 2016-09, Vol.33, p.47-50 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Abstract Operating room surgical table failure is a rare event but can lead to a dangerous situation when it does occur. The dangers can be compounded in the presence of obesity, especially in the anesthetized or sedated patient. We present a case of a near-miss fall of a morbidly obese patient while turning the patient in preparation to transfer from the operating room table to the hospital bed when 2 fractured bolts in the tilt cylinder mechanism led to an operating room table failure. |
---|---|
ISSN: | 0952-8180 1873-4529 |
DOI: | 10.1016/j.jclinane.2016.01.016 |