Accuracy of Coronary CT Angiography Using a Submillisievert Fraction of Radiation Exposure: Comparison With Invasive Coronary Angiography

Coronary computed tomography angiography (CTA) is increasingly being used for evaluation of coronary artery disease (CAD). As a result of the widely reported potential of carcinogenic risk from x-ray based examinations, many strategies have been developed for dose reduction with CTA. The purpose of...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of the American College of Cardiology 2014-08, Vol.64 (8), p.772-780
Hauptverfasser: STEHLI, Julia, FUCHS, Tobias A, BULL, Sacha, CLERC, Olivier F, POSSNER, Mathias, BUECHEL, Ronny R, GAEMPERLI, Oliver, KAUFMANN, Philipp A
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Coronary computed tomography angiography (CTA) is increasingly being used for evaluation of coronary artery disease (CAD). As a result of the widely reported potential of carcinogenic risk from x-ray based examinations, many strategies have been developed for dose reduction with CTA. The purpose of this study was to assess the diagnostic accuracy of CTA acquired with a submillisievert fraction of effective radiation dose reconstructed with a model-based iterative reconstruction (MBIR) using invasive coronary angiography (ICA) as a standard of reference. In 36 patients (body mass index range 17 to 39 kg/m(2)) undergoing ICA for CAD evaluation, a CTA was acquired using very low tube voltage (80 to 100 kV) and current (150 to 210 mA) and was reconstructed with MBIR. CAD (defined as ≥50% luminal narrowing) was assessed on CTA and on ICA. CTA resulted in an estimated radiation dose exposure of 0.29 ± 0.12 mSv (range 0.16 to 0.53 mSv), yielding 96.9% (436 of 450) interpretable segments. On an intention-to-diagnose basis, no segment was excluded, and vessels with at least 1 nonevaluable segment and no further finding were classified as false positive. This resulted in a sensitivity, specificity, positive, and negative predictive value and accuracy of 100%, 74%, 77%, 100%, and 86% per patient and 85%, 86%, 56%, 96%, and 85% per vessel, respectively. The use of MBIR reconstruction allows accurate noninvasive diagnosis of CAD with CTA at a submillisievert fraction of effective radiation dose comparable with a chest x-ray in 2 views.
ISSN:0735-1097
1558-3597
DOI:10.1016/j.jacc.2014.04.079