Variability of FeNO in healthy subjects at 2240 meters above sea level
Fraction of exhaled nitric oxide (FeNO) is a marker of eosinophilic airway inflammation. Altitude above sea level can affect measurements of this index, but there is only limited information regarding the diurnal variation (ante meridiem vs. post meridiem) and reproducibility of FeNO on consecutive...
Gespeichert in:
Veröffentlicht in: | Journal of clinical monitoring and computing 2016-08, Vol.30 (4), p.445-449 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Fraction of exhaled nitric oxide (FeNO) is a marker of eosinophilic airway inflammation. Altitude above sea level can affect measurements of this index, but there is only limited information regarding the diurnal variation (ante meridiem vs. post meridiem) and reproducibility of FeNO on consecutive days at moderate altitudes. To evaluate the diurnal variability of FeNO and assess its reproducibility over five consecutive days in healthy individuals living at 2240 m, and to compare the FeNO readings taken with two different analyzers. Healthy non-smoking adults were measured using NIOX MINO
®
or NOA 280i
®
devices. One group (n = 10) had readings taken morning and afternoon for five consecutive days with the NIOX MINO
®
equipment; while the second group (n = 17) was measured on only one morning but by both the electrochemical analyzer (NIOX MINO
®
) and the chemiluminescence method (NOA 280i
®
). The study group consisted of 27 subjects aged 28.7 ± 6 years. Morning and afternoon FeNO measurements were 15.2 ± 7.5 ppb and 15.2 ± 7.9 ppb (
p
= 0.9), respectively. The coefficient of variation (CV) of these measurements (a.m. vs. p.m.) was 10.7 %, and the coefficient of repeatability (CR), 4.2 ppb. The concordance correlation coefficient (CCC) between the two measures (morning vs. afternoon) was 0.91. The CV and CR of the five morning readings were 15.4 % and 4.3 ppb, respectively; while those of the five afternoon measures were 13.6 % and 3.5 ppb, respectively. The CCC between the NIOX MINO
®
equipment and the NOA-280i
®
device was 0.8, with 95 % limits of agreement of −8.35 to 0.29 ppb. In adults living at 2240 m above sea level, FeNO measurements show minimal diurnal variation, and readings are reproducible ( |
---|---|
ISSN: | 1387-1307 1573-2614 |
DOI: | 10.1007/s10877-015-9737-0 |