Ex vivo estimation of cementless acetabular cup stability using an impact hammer
Highlights • Acetabular cup implants were inserted into 20 bovine bone samples with different cavity diameter. • An impact hammer was designed and was used to measure the impact momentum. • Tangential stability tests were performed. • The AC implant stability and the impact momentum were significant...
Gespeichert in:
Veröffentlicht in: | Medical engineering & physics 2016-02, Vol.38 (2), p.80-86 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Highlights • Acetabular cup implants were inserted into 20 bovine bone samples with different cavity diameter. • An impact hammer was designed and was used to measure the impact momentum. • Tangential stability tests were performed. • The AC implant stability and the impact momentum were significantly correlated (R2 = 0.83). |
---|---|