Extraction of Benzoylecgonine from Urine Specimens with Cerex Polycrom™ Clin II Solid-Phase Extraction Columns and the Speedisk™ Pressure Processor
Benzoylecgonine (BZE) extraction from urine was explored using Cerex Polycrom Clin II solid-phase extraction (SPE) columns and the Speedisk 48 Pressure Processor as an alternative to the Prep™1 automated sample processor and XTRX™ Type RP/W columns. Linearity for urine standards extracted using the...
Gespeichert in:
Veröffentlicht in: | Journal of analytical toxicology 2001-10, Vol.25 (7), p.616-620 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Benzoylecgonine (BZE) extraction from urine was explored using Cerex Polycrom Clin II solid-phase extraction (SPE) columns and the Speedisk 48 Pressure Processor as an alternative to the Prep™1 automated sample processor and XTRX™ Type RP/W columns. Linearity for urine standards extracted using the Cerex-Speedisk method ranged from 20 to 3000 ng/mL. The mean recovery at the 100-ng/mL cutoff for three lots of columns was 92%. The mean of the within-run means for three batches, which had coefficients of variations of 1.8% or less, was 101.3 ng/mL at the lO0-ng/mL cutoff level. Forty-six specimens known to contain BZE were analyzed by both the Prep1-Type RP/W and Cerex-Speedisk methods. The correlation for specimen BZE concentrations between the two methods gave an r2 of 0.9999 and a slope of 1.03. The Cerex-Speedisk system is an inexpensive alternative to the Prep1-Type RP/W system. It is less costly, requires little maintenance, has a small footprint, is hood compatible, and can process four times the number of specimens in a given time. |
---|---|
ISSN: | 0146-4760 1945-2403 |
DOI: | 10.1093/jat/25.7.616 |