CALIBRATION OF THERMOLUMINESCENCE AND FILM DOSEMETERS FOR SKIN DOSES FROM HIGH-ACTIVITY MICROPARTICLES
The use of EXT-RAD™ extremity TLDs and radiochromic film to measure doses from primarily beta-emitting microparticles is discussed. Specific calibration techniques have been developed, using both Monte Carlo modelling and experiments. Results for a (90)Sr/(90)Y microparticle are presented to illustr...
Gespeichert in:
Veröffentlicht in: | Radiation protection dosimetry 2016-09, Vol.170 (1-4), p.173-176 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The use of EXT-RAD™ extremity TLDs and radiochromic film to measure doses from primarily beta-emitting microparticles is discussed. Specific calibration techniques have been developed, using both Monte Carlo modelling and experiments. Results for a (90)Sr/(90)Y microparticle are presented to illustrate the general techniques and to demonstrate reasonable agreement between the dosimetry methods. |
---|---|
ISSN: | 0144-8420 1742-3406 |
DOI: | 10.1093/rpd/ncv437 |