Type Testing of Model 7200 Automatic TLD Reader
The type testing of measuring devices is one of the most important parts of a quality management system in a personal dosimetry services program. In this study, based upon the International Electrotechnical Commission (IEC) 62387 criteria, a reader-testing program was performed for a home-made perso...
Gespeichert in:
Veröffentlicht in: | Radiation protection dosimetry 2017-04, Vol.174 (1), p.68-73 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The type testing of measuring devices is one of the most important parts of a quality management system in a personal dosimetry services program. In this study, based upon the International Electrotechnical Commission (IEC) 62387 criteria, a reader-testing program was performed for a home-made personal thermoluminescent dosimetry (TLD) reader. The stability of the reader, the effects of light exposure, temperature and fluctuations of primary power supply on TLD read-outs as the main parameters were investigated in this program. Moreover, this study assesses some important criteria of dosimetry system including the non-linearity of response, reusability, after effect and overload that may include significant contribution in the performance of a reader. The results showed that the TLD reader met all requirements of the IEC for the reader tests by a large margin. |
---|---|
ISSN: | 0144-8420 1742-3406 |
DOI: | 10.1093/rpd/ncw100 |